Chapter 6: Phase 3 - Protocols
Measurement protocols and clinical assessment procedures
6.1 TMS-EEG Measurement Protocol
Equipment Requirements
- TMS System: Navigated TMS with biphasic pulse capability (Nexstim, Magstim, MagVenture)
- EEG System: ≥60 channels, ≥5000 Hz sampling, TMS-compatible amplifiers
- Neuronavigation: MRI-based head positioning
- Ancillary: EMG for motor threshold, ECG for artifact removal
Standard PCI Protocol Steps
- 1. Subject Preparation: Position comfortably, apply EEG cap, verify impedances <5 kΩ
- 2. Neuronavigation: Co-register head position with structural MRI
- 3. Target Selection: Identify left premotor cortex (BA6) as primary target
- 4. Baseline Recording: 5 minutes resting EEG with fixation cross
- 5. TMS Stimulation: 200 biphasic pulses at 0.5-1 Hz, 80-120% motor threshold
- 6. EEG Recording: 500ms pre- and post-stimulus epochs
- 7. Quality Check: Verify ≥150 artifact-free epochs
TMS Parameters
| Parameter | Value |
| Target Region | Left premotor cortex (BA6) |
| Intensity | 80-120% resting motor threshold |
| Pulse Type | Biphasic |
| Frequency | 0.5-1.0 Hz (randomized ISI) |
| Pulse Count | 200-300 pulses |
| Recording Window | -500 to +500 ms |
6.2 PCI Computation Pipeline
1. PREPROCESSING
├── Artifact Removal
│ ├── TMS artifact: interpolation (-2 to 15 ms)
│ ├── Muscle: ICA/SSP
│ ├── Eye: ICA regression
│ └── Cardiac: template subtraction
├── Filtering: 1-80 Hz bandpass
└── Epoching: -100 to 500 ms
2. BINARIZATION
├── Source localization
├── Threshold: 3.5 SD from baseline
└── Time window: 20-300 ms post-TMS
3. COMPRESSION
├── Create spatiotemporal matrix
├── Flatten to 1D string
└── Apply Lempel-Ziv algorithm
4. NORMALIZATION
└── PCI = L(SS) / (s × t × H(p))
6.3 Clinical Assessment Protocol
Disorders of Consciousness Evaluation
Important: PCI should complement, not replace, behavioral assessment (CRS-R).
Standard evaluation workflow:
- Conduct Coma Recovery Scale-Revised (CRS-R) - 5 sessions over 2 weeks
- Perform TMS-EEG during maximal arousal
- Calculate PCI and classify state
- Compare behavioral and electrophysiological findings
- Consider additional tests if discrepancy exists
PCI Classification Thresholds
| PCI Value | Classification | Clinical Interpretation |
| < 0.20 | VS/UWS (likely) | Very low complexity, consistent with deep unconsciousness |
| 0.20 - 0.31 | Borderline | Below threshold, repeat testing recommended |
| 0.31 - 0.45 | MCS | Above threshold, evidence of consciousness |
| > 0.45 | Conscious | High complexity, consider LIS if unresponsive |
6.4 Anesthesia Monitoring Protocol
PRE-INDUCTION
└── Baseline PCI: 0.45-0.65 expected
INDUCTION
└── Monitor for PCI drop
MAINTENANCE
├── Target PCI: 0.15-0.25
└── Alert if PCI > 0.30 (possible awareness)
EMERGENCE
└── Recovery when PCI > 0.35
Drug-Specific Effects
| Agent | Typical PCI | Notes |
| Propofol | 0.10-0.20 | Linear dose-response |
| Sevoflurane | 0.15-0.25 | Sigmoid dose-response |
| Ketamine | 0.30-0.45 | Preserved complexity (dissociative) |
| Dexmedetomidine | 0.25-0.35 | Sleep-like, rousable |
6.5 Quality Standards
Quality Criteria:
- Impedances: <5 kΩ all channels
- Artifact-free epochs: ≥150
- Signal amplitude: <500 μV
- Test-retest reliability: ICC >0.7
6.6 Interpretation Guidelines
Important caveats:
- Single measurement should not determine diagnosis alone
- Medications (especially sedatives) can affect PCI
- Repeat measurements recommended across multiple sessions
- Clinical behavioral assessment remains essential
- PCI measures capacity for consciousness, not content
For borderline cases (PCI within 0.05 of threshold):
- Repeat measurement 3x over 1 week
- Report range and trend
- Do not assign definitive classification
- Recommend continued monitoring